J-2800 PFE TESTER FOR SURGICAL MASK PARTICLE FILTRATION EFFICIENCY

EQUIPPED WITH SALT AEROSOL GENERATOR WHICH PRODUCES AEROSOLS WITH SPECIFIC PARTICLE SIZE AND CONCENTRATION

SKU: J-2800 Category:
 

Description

EQUIPPED WITH SALT AEROSOL GENERATOR WHICH PRODUCES AEROSOLS WITH SPECIFIC PARTICLE SIZE AND CONCENTRATION
INCLUDES A RANGE OF SPECIALIZED FIXTURES FOR TESTING OF MULTIPLE MASK TYPES
LONG-LIFE PHOTOMETER MODULE FOR SAMPLING TIME MEASUREMENT AND PROMPT LIGHT CLEANING
AEROSOL GENERATOR AUTOMATICALLY CONTROLLED
AUTOMATICALLY CALUCLATES CAPTURE EFFICIENCY AND MASK AIRFLOW RESISTANCE
BUILT IN ELECTRONIC FLOWMETER AND SAMPLING PUMP ARE CONSTRUCTED FOR LONG LIFE STABILITY
TEST FLOW PARAMETER RANGE 5 TO 100 LITERS PER MINUTE, RESOLUTION 5 TO 100 LITERS PER MINUTE WITH MAXIMUM ERROR RATE 1%
PRESSURE DETECTION RANGE: 0 TO 2500 Pa AT RESOLUTION 0.1 Pa WITH MAXIMUM ERROR 1%
CONCENTRATION DETECTION RANGE : (0.001 TO 100) 0 grams PER LITER
DETECTION ACCURACY: 1% OF THE READ VALUE IN THE RANGE OF 0.01% TO 100%
DETECTION REPEATABILITY: 5% OF THE READ VALUE IN THE RANGE OF 0.01% TO 100%
MEDIAN DIAMETER OF AEROSOLIZED SALT PARTICLES (CMD) (0.075±0.02)
MEDIAN AEROSOL PARTICLE NUMBER DIAMETER (CMD) (OPTIONAL) :(0.185±0.02) M
OPERATING TEMPERATURE: 20±5℃
INSTRUMENT NOISE: <60DB(A)
POWER SUPPLY: AC220V±10%, 50Hz (OR AS PER CUSTOMER)
DIMENSIONS: L 700 x W 720 x H 1450mm

QUANTITY: 1