A-2611 ULTRAVIOLET AGING TESTING CHAMBER BOX TYPE

WORKING CHAMBER DIMENSIONS: 1130 mm × 400 mm × 500 mm
EXTERNAL SIZE: 1300 mm × 1400 mm × 700 mm
LAMP DISTANCE TO PRODUCT: ADJUSTABLE (MAX 450 mm)
THE SHORTEST PARALLEL SURFACE DISTANCE BETWEEN THE SPECIMEN AND THE LAMP
SURFACE: ABOUT 100 mm
WAVELENGTH RANGE: UV-A, WAVELENGTH RANGE 315 TO 400 NM / UV-B, WAVELENGTH
RANGE 290 TO 315 NM
RADIATION INTENSITY: 1.2 W/M2/340 NM
TEMPERATURE RESOLUTION: 0.1 ℃

SKU: A-2611 Category:
 

Description

FLUORESCENT UV LAMP IS IMPORTED FOR THIS ACCELERATED AGING TEST CHAMBER. CAN SIMULATE HARM BY SUNLIGHT, RAINWATER AND DEW. UV WEATHER RESISTANCE CHAMBER USE FLUORESCENT UV LAMP TO SIMULATES THE EFFECT OF SUNLIGHT, CONDENSATION MOISTURE IS USED TO SIMULATE DEW. THE TESTED MATERIAL IS PLACED IN A LOOP PROGRAM THAT ALTERNATES LIGHT AND MOISTURE AT A CERTAIN TEMPERATURE. THE MATERIAL WAS TESTED FOR ACCELERATED WEATHERABILITY. UV CHAMBER CAN RE-CREATE HAZARDS THAT OCCUR OUTDOORS FOR MONTHS OR YEARS.THE TYPES OF HAZARDS INCLUDE: DISCOLORATION, CHALKINESS, POWDER LIGHT, CRACKING, TURBIDITY, BUBBLES, BRITTLE, STRENGTH, DECAY, OXIDATION, ETC. THIS MACHINE INCLUDES SPRAY DEVICE.

CAN SIMULATE ULTRAVIOLET, RAIN, HIGH TEMPERATURE, HIGH HUMIDITY, CONDENSATION, DARKNESS IN NATURAL ENVIRONMENT FOR THIS ACCELERATED AGING TEST CHAMBER.

BY REPRODUCING THESE CONDITIONS, MERGE INTO A LOOP AND LET IT AUTOMATICALLY EXECUTE THE NUMBER OF COMPLETED LOOPS. THIS IS HOW THE UV AGING TEST BOX WORKS. IN THIS PROCESS, THE EQUIPMENT CAN MONITOR BLACKBOARD TEMPERATURE AND SINK TEMPERATURE AUTOMATICALLY; CONTROL THE DEVICE BY CONFIGURING RADIATION MEASUREMENT (OPTIONAL), IRRADIANCE CAN BE MEASURED AND CONTROLLED, STABILIZE THE IRRADIANCE AT 0.76 W / M2 /340 NM OR A SET VALUE, GREATLY EXTEND THE SERVICE LIFE OF THE LAMP

SPECIFICATIONS:
WORKING CHAMBER DIMENSIONS: 1130 mm × 400 mm × 500 mm
EXTERNAL SIZE: 1300 mm × 1400 mm × 700 mm
LAMP DISTANCE TO PRODUCT: ADJUSTABLE (MAX 450 mm)
THE SHORTEST PARALLEL SURFACE DISTANCE BETWEEN THE SPECIMEN AND THE LAMP
SURFACE: ABOUT 100 mm
WAVELENGTH RANGE: UV-A, WAVELENGTH RANGE 315 TO 400 NM / UV-B, WAVELENGTH
RANGE 290 TO 315 NM
RADIATION INTENSITY: 1.2 W/M2/340 NM
TEMPERATURE RESOLUTION: 0.1 ℃
LUMINOUS TEMPERATURE RANGE: 50 ℃ TO 70 ℃ / TEMPERATURE TOLERANCE ±3 ℃
CONDENSATION TEMPERATURE RANGE: 40℃ TO 60℃/TEMPERATURE TOLERANCE ±3 ℃
BLACKBOARD THERMOMETER MEASURING RANGE: 30 TO 80 ℃/TOLERANCE ±1 ℃
TEMPERATURE CONTROL TYPE: PID SELF SETTING TEMPERATURE CONTROL MODE
HUMIDITY RANGE: ABOUT 45% TO 70% R.H (ILLUMINATION CONDITION / 98%
(CONDENSATION STATE)
SINK REQUIREMENTS: WATER DEPTH IS NOT GREATER THAN 25 mm AND THERE IS AN AUTOMATIC WATER SUPPLY CONTROLLER
STANDARD SPECIMEN SIZE: 1000 × 400 mm
RECOMMENDED ENVIRONMENT FOR INSTRUMENT USE: 5 TO 35 ℃, 40% TO 85% R·H, DISTANCE TO THE WALL IS 300 mm

CHAMBER STRUCTURE:
SPRAY HEADER
BOX DOOR
UV LAMP
FAN

COMPLY WITH INTERNATIONAL STANDARDS
ASTM G 153, ASTM G 154, ASTM D 4329, ASTM D 4799, ASTM D 4587, SAE, J2 020, ISO 4892 ALL CURRENT UV AGING TEST STANDARDS

SUBENTRY INTRODUCE
ILLUMINANT – THE ILLUMINANT INCORPORATES 8 PCS RATED IMPORTED UV FLUORESCENT LAMP AS A LIGHT SOURCE WHICH POWER IS 40 W. UV FLUORESCENT LAMP TUBE, DISTRIBUTED ON BOTH SIDES OF THE MACHINE. 4 PCS ON EACH SIDE. UVA-340 AND UVB-313 LIGHT SOURCE IS CONFIGURABLE BY THE USER. UVA-340 LUMINOUS SPECTRAL OF THE LAMP IS MAINLY CONCENTRATED AROUND 340 NM WAVE-LENGTH, UVB-313 LUMINOUS SPECTRAL OF THE LAMP IS MAINLY CONCENTRATED AROUND WAVE-LENGTH 313 NM. UVA-340 FLUORESCENT LIGHT ENERGY OUTPUT DECLINE WITH TIME,IN ORDER TO REDUCE THE TEST EFFECT CAUSED BY LIGHT ENERGY ATTENUATION,REPLACE AN OLD LAMP WITH A NEW FOR 8 LAMP TUBES OF THE FLUORESCENT LAMP EVERY 1/4 LIFE TIME, ULTRAVIOLET LIGHT SOURCE IS ALWAYS MADE UP OF NEW AND OLD LIGHTS, SO YOU CAN GET A CONSTANT OUTPUT OF LIGHT ENERGY. THE EFFECTIVE SERVICE LIFE OF THE LAMP TUBE CAN BE ABOUT 1600 HOURS
• BLACKBOARD TEMPERATURE, CONDENSING TEMPERATURE IS CONTROLLED BY THE CONTROLLER
• THE OTHERS ARE IMPORTED ELECTRONIC COMPONENTS IRRADIANCE UNIFORMITY: ≤4 %(AT THE SURFACE OF THE SAMPLE)
POWER CONTROL – BLACKBOARD TEMPERATURE MONITORING: STANDARDIZED PT-100 BLACKBOARD TEMPERATURE SENSOR. ACCURATE CONTROL OF THE SAMPLE SURFACE TEMPERATURE DURING THE TEST PROCESS. BLACKBOARD TEMPERATURE SETTING RANGE: BPT 40 TO 75℃: ACTUAL MAXIMUM TEMPERATURE OF PROTECTION DEVICE INSIDE THE MACHINE IS 93 ℃ ± 10%.
BLACKBOARD TEMPERATURE CONTROL ACCURACY: ±0.5 ℃
• TANK TEMPERATURE MONITORING: ONE TEST SECTION IS THE DARK CONDENSATION PROCESS IN THE LOOP TEST PROCESS, IT REQUIRES THE BOX TO BE ABLE TO PRODUCE HIGH TEMPERATURE SATURATED WATER VAPOR, WHEN THE STEAM HITS A RELATIVELY COLD SURFACE OF THE SAMPLE. DEW WILL CONDENSE ON THE SURFACE OF THE SAMPLE. THE TANK IS LOCATED AT THE BOTTOM OF THE CHAMBER WITH BUILT-IN ELECTRIC HEATER. TANK TEMPERATURE CONTROL RANGE: 40 TO 60 ℃
• THE TEST CHAMBER IS EQUIPPED WITH A TIME CONTROLLER, RANGE 0 TO 530 H, BACKUP DATA FOR POWER FAIL
• SAFETY PROTECTION DEVICE
TEMPERATURE PROTECTION IN THE CHAMBER: WHEN THE TEMPERATURE IN THE CHAMBER EXCEEDS 93 ℃ ± 10%, MACHINE WILL AUTOMATICALLY CUT OFF THE SUPPLY OF THE LAMP AND HEATER, GO INTO EQUILIBRIUM STAGE AND COOL
LOW WATER TANK ALARM, PREVENT HEATER FROM BURNING

STANDARD SAMPLE HOLDER
CHAMBER MATERIAL
• INNER MATERIAL USE SUS304# STAINLESS STEEL PLATE
• OUTER SHELL USE SUS304# STAINLESS STEEL PLATE
• THE SAMPLE FRAME IS MADE OF STAINLESS STEEL AND ALUMINUM ALLOY FOR EASY ACCESS
WHOLE MACHINE PROFILE
• DIMENSION: ABOUT 1600 mm × 1200 mm × 700 mm
• WEIGHT: ABOUT 150 KG
MAINFRAME WORKING ENVIRONMENT
• POWER REQUIREMENT: 220 VOLTZ AND 50/60 Hz
• ENVIRONMENT: 5 TO 35 ℃,0 TO 80% RH,EXCELLENT VENTILATION, INDOOR ENVIRONMENT CLEANING
• AREA OF WORK: ABOUT 234 × 353 CM
• DRAINING: A DRAIN DITCH IS NEEDED ON THE FLOOR NEAR THE MAIN ENGINE
• FOR EASE OF MOVEMENT, FOOT WHEEL IS INSTALLED AT THE BOTTOM OF THE INSTRUMENT, POSITION OF THE TEST MACHINE IS FIXED WITH A U-RING

CONTROL INSTRUMENT – THE DEVICE USE A TRUE COLOR TOUCH SCREEN, PID TEMPERATURE INTELLIGENT CONTROL DEVICE, HIGH PRECISION TEMPERATURE CONTROL AND GOOD STABILITY

INSTALLATION SITE – FOR THE CONVENIENCE OF HEAT DISSIPATION AND MAINTENANCE.INSTALLATION OF THE EQUIPMENT MUST MEET THE FOLLOWING CONDITIONS:
• ADEQUATE MAINTENANCE SPACE SHOULD BE MAINTAINED BETWEEN THE EQUIPMENT AND ADJACENT WALLS OR OTHER EQUIPMENT AFTER INSTALLATION
• IN ORDER TO STABILIZE THE FUNCTION OF TEST EQUIPMENT, PERFORMANCE,CHOOSE TEMPERATURE IS 15 ℃ TO 28 ℃, RELATIVE HUMIDITY IS NOT GREATER THAN 85% SITE
• ENVIRONMENT OF THE INSTALLATION SITE. TEMPERATURE MUST NOT CHANGE SHARPLY ALL YEAR
• IT SHOULD BE INSTALLED ON A LEVEL GROUND(THE LEVEL SHALL BE DETERMINED ON THE GROUND BY A LEVEL METER DURING INSTALLATION)
• IT SHOULD BE INSTALLED IN PLACES WITHOUT DIRECT SUNLIGHT
• IT SHOULD BE INSTALLED IN A WELL-VENTILATED PLACE
• SHOULD BE INSTALLED AWAY FROM COMBUSTIBLE MATERIALS, EXPLOSIVE, HIGH TEMPERATURE SOURCE
• IT SHOULD BE INSTALLED IN A LITTLE DUSTY PLACE 9. INSTALL AS CLOSE TO THE POWER SOURCE AS POSSIBLE

STRUCTURE PROFILE – THE EQUIPMENT SHELL AND INNER ROOM ARE MADE OF HIGH QUALITY SUS304 STAINLESS STEEL MATERIAL. THE MOST ADVANCED PROCESSING EQUIPMENT IN CHINA IS USED FOR PROCESSING AND FORMING. THE SURFACE OF THE SHELL IS SPRAYED WITH PLASTIC, BEAUTIFUL, SMOOTH, COLOR COORDINATION, SMOOTH SURFACE. THE DEFECTS OF SURFACE DAMAGE CAUSED BY ALUMINUM PLATE SURFACE SPRAYING ARE AVOIDED. THE EQUIPMENT WORKSHOP IS EQUIPPED WITH A HUMIDIFYING HEATER, LIQUID LEVEL SWITCH, BLACKBOARD TEMPERATURE SENSOR, ETC. THERE ARE ALSO EIGHT ULTRAVIOLET RADIATION FLUORESCENT AND ULTRAVIOLET RADIATION METERS (OPTIONAL). HIGH QUALITY IMMOVABLE PU WHEEL IS INSTALLED AT THE BOTTOM OF THE MACHINE. THE MACHINE CAN BE MOVED TO THE SPECIFIED LOCATION EASILY. FINALLY, THE CASTER IS FIXED
• THE EQUIPMENT IS RUNNING AND BE SURE TO KEEP ENOUGH WATER
• THE OPENING TIME OF THE DOOR SHOULD BE MINIMIZED DURING THE TEST PERIOD